xrayCase Study #23

That is incorrect. The correct answer is that there was insufficient vertical angulation used when this film was exposed. Consequently, the teeth are elongated. This radiograph cannot be used for diagnostic purposes as it does not represent the true length of the tooth, the apices are missing and the image is disorted.

To make this film diagnostic, the vertical angulation needs to be increased.

Next Case